PART VIIMaintenance and support
CHAPTER 25

Maintenance routine

The Aquarino KH asks little, but it asks regularly. This routine keeps measurements trustworthy over the years.

FrequencyWhat to do
WeeklyLook at the last 24 hours in the graph · check the aeration foam in both chambers · check the buffer reservoir level
MonthlyRun a chemical test and compare with the equipment · partial calibration if needed · inspect the chambers visually
QuarterlyClean the chambers internally, then run a full calibration · check the doser tubing
YearlyReplace the pH probe and log the replacement · replace the doser's internal tubing

25.1 Cleaning the chambers

  1. Empty the reservoirs completely under Configuration → Reservoirs → Empty.
  2. Open the lid and clean the inside of the chambers.
  3. Close it, run Fill and perform a full calibration (Chapter 10).

There is a video of the procedure at https://youtube.com/shorts/qm--dZ7LNUg

25.2 Memory card

The card holds the history, the event log and the interface files. When it fails, the symptoms are characteristic: graphs stop appearing, updates do not complete, or System Status shows SD: Not Detected.

The fix is to format or replace the card. Talk to support first — in many cases we can identify the cause remotely.

25.3 pH probe

  • Rinse only with drinking water — never with reverse-osmosis, distilled or deionised water, which attack the bulb. Never wipe the bulb with cloth or paper.
  • Never let the probe dry out — more than 30 minutes in the air can damage it permanently.
  • Replace it every 12 months, or sooner if Probe Health falls below 55%.
  • Always log the change under Configuration → pH Probe → Log probe replacement.