Aquarino KH — User Manual · Printed from manual.aquarino.com.br/en/kh/4-the-ph-probe-installation-and-calibration/

PART IInstall
CHAPTER 4

The pH probe: installation and calibration

4.1 Installing the pH probe

The pH probe is delicate and needs careful handling. Set aside a few minutes and follow the steps calmly.

  1. Remove the silicone protective cap from the probe only when you are about to calibrate, and rinse the tip in a cup of drinking water.
  2. Connect the probe's BNC to the side input marked PH-IN.
  3. On first installation, calibrate the probe before placing it in the compartment (section 4.2). The equipment arrives empty, which is the right condition for calibrating.
  4. Once it is calibrated, carefully insert the probe into the measurement compartment of the Aquarino KH.
Step 1 — remove the protective cap and rinse the probe.
Step 1 — remove the protective cap and rinse the probe.
Step 2 — the probe's BNC goes into the PH-IN input, the round one at the top. KH-OUT, just below it, is the output for another controller.
Step 2 — the probe's BNC goes into the PH-IN input, the round one at the top. KH-OUT, just below it, is the output for another controller.
Step 4 — once calibrated, insert the probe into the measurement compartment.
Step 4 — once calibrated, insert the probe into the measurement compartment.

4.2 Two-point calibration

This screen brings together everything about the probe: the current reading, the health index, the installation date, probe care and the calibration wizard. You reach it through Configuration → pH Probe.

The pH Probe screen: reading, the Calibrate probe button, health, installed probe and probe care.
The pH Probe screen: reading, the Calibrate probe button, health, installed probe and probe care.

Probe calibration is what lets the equipment display the real pH of your aquarium. It is not required in order to measure KH — see section 25.3.

Tap Calibrate probe. A wizard opens and takes you through four steps, always in the same order. Set aside about 50 minutes: each point needs 20 minutes to stabilise.

StepWhat happens
1 · PrepareThe wizard reminds you what to have ready and the main thing to watch with the probe
2 · Low pointThe probe in the pH 7.00 solution, 20 minutes of stabilisation and the first point saved
3 · High pointThe same with the pH 10.00 solution
4 · DoneConfirmation that the probe is calibrated, and its slope
Step 1, Prepare, and step 2, Low point, during the countdown: the ring on Stabilizing and the calibrate button still greyed out.
Step 1, Prepare, and step 2, Low point, during the countdown: the ring on Stabilizing and the calibrate button still greyed out.

Step 1 · Prepare

Before you start, have ready:

  • The equipment empty, under Configuration → Reservoirs → Empty, with the probe out of the compartment. On first installation it already arrives empty.
  • Two fresh calibration solutions: pH 7.00 and pH 10.00 — the ones supplied with the product.
  • One container for each solution.
  • Clean drinking water to rinse the probe.

The point values are fixed: the wizard always uses pH 7.00 for the low point and pH 10.00 for the high point. Tap Next.

Steps 2 and 3 · Low point and high point

Both points follow the same routine; only the solution changes:

  1. Rinse the probe in drinking water and dry it with soft paper, without rubbing the bulb.
  2. Place the probe straight into the solution container — pH 7.00 for the low point, pH 10.00 for the high point — without touching the bulb to the bottom.
  3. Tap Probe in place — start stabilization. A 20-minute countdown starts on a ring that fills up as it goes: Stabilizing, and at the end Stable.
  4. Keep the probe still until the countdown ends — that is what the line under the ring reminds you while it runs: Keep the probe still until the countdown ends. When the countdown finishes, the line changes to Reading stable — you can calibrate the point.
  5. When Stable reading appears, tap Calibrate Low Point — or Calibrate High Point. The wizard moves on to the next step by itself.
On screenWhat it means
Calibrate button greyed outThe countdown has not finished yet. It is only enabled after the 20 minutes — there is no skipping it
Restart countdownShown during the countdown. Use it if the probe moved or was disturbed: the countdown goes back to 20 minutes
Sent to the deviceAfter the low point: the command reached the equipment
Saved and confirmed by the deviceAfter the high point: the equipment confirmed that the probe is calibrated
Step 3, High point, with the countdown over: Stable reading and the button enabled. Beside it, step 4, Done, with the probe slope.
Step 3, High point, with the countdown over: Stable reading and the button enabled. Beside it, step 4, Done, with the probe slope.

Step 4 · Done

The wizard shows pH probe calibrated. Wash the probe in drinking water and put it back in the measurement compartment. The same screen shows the probe's slope — a score of its sensitivity, as a percentage of the ideal:

SlopeRangeWhat to do
90 to 105GoodProbe in good condition
80 to 89AttentionProbe ageing — plan the replacement
below 80ReplaceLow slope — replace the probe

Tap Finish to close the wizard. The screen then shows Probe calibrated.

The order of every probe calibration: empty, calibrate, put the probe back, fill and enter the KH.
The order of every probe calibration: empty, calibrate, put the probe back, fill and enter the KH.

4.3 Probe care

The Probe care panel, on the same screen, sums up the three rules that do the most to extend the probe's life:

CareWhy
Every 3 months, calibrate with the pH 7.00 and 10.00 solutionsThe probe drifts over time; recalibrating keeps the displayed pH reliable. Do it alongside the quarterly cleaning, which already empties the equipment (Chapter 13)
No reverse osmosis: never use it to rinse or store the probeWater without salts attacks the bulb and the internal electrolyte
Never dry: keep the bulb moist at all timesA bulb that dries out loses sensitivity and may not recover