Aquarino KH — User Manual · Printed from manual.aquarino.com.br/en/kh/25-how-the-aquarino-kh-measures/

PART VReference
CHAPTER 25

How the Aquarino KH measures

25.1 Measuring without reagents

The Aquarino KH uses a patented method to track the changes in your alkalinity, taking as its basis a reference alkalinity that you provide beforehand.

In practice this has two consequences that are worth half of this manual:

  • Since no reagent is consumed, measuring costs nothing — you can measure every hour, every day, indefinitely.
  • Since the method starts from a reference you provide, measurement quality depends directly on the quality of that information. That is why calibration gets a chapter of its own and is the most important step of the installation.

25.2 The measurement cycle

Each measurement takes about 30 minutes, and you follow the progress live, as a percentage and a time remaining, on the home screen.

StageWhat happens
PreparationThe internal chambers are emptied and rinsed
CollectionA fresh sample is drawn from your aquarium
AerationThe chambers are aerated until they stabilise
ReadingThe probe takes the required readings
ResultKH is calculated, recorded and displayed
Close-outAlerts are sent and the correction for the period is decided

Throughout the cycle the equipment keeps responding normally in the app, and your calcium reactor carries on with the plan set at the previous measurement — nothing sits idle waiting for the result.

25.3 The pH probe: what actually matters

The pH probe is the most delicate part of the set, and the one that most affects confidence in the readings. Two things are commonly confused:

What it is forNeeded to measure KH?
Probe calibration (pH 7 and pH 10)Displaying the real pH of your aquarium on screen and in the graphNo
Probe health (a 0 to 100% index)Telling you whether the probe is still reliable — it considers signal noise, slope, offset and probe ageYes, this is what counts

In other words: you could never calibrate the probe and KH would still be measured. Without calibration, only the Tank pH field stays empty and the pH series is not recorded in the graph.