Aquarino KH — User Manual · Printed from manual.aquarino.com.br/en/kh/13-maintenance-routine/

PART IIDay-to-day use
CHAPTER 13

Maintenance routine

The Aquarino KH asks little, but it asks regularly. This routine keeps measurements trustworthy over the years.

FrequencyWhat to do
WeeklyLook at the last 24 hours in the graph · check the aeration foam in both chambers · check the buffer reservoir level
MonthlyRun a chemical test and compare with the equipment · partial calibration if needed · inspect the chambers visually
QuarterlyClean the chambers internally and calibrate the pH probe with the equipment empty, followed by Fill and the reference alkalinity (section 13.1) · check the doser tubing
YearlyReplace the pH probe, log the replacement and only then calibrate the new one, with the equipment empty · replace the internal tubing of the dosing pumps

13.1 Cleaning the chambers

  1. Empty the reservoirs completely under Configuration → Reservoirs → Empty.
  2. Open the lid and clean the inside of the chambers.
  3. With the equipment still empty, calibrate the pH probe with the wizard in section 4.2: no measurement starts and no motor runs during the calibration.
  4. Put the probe in the compartment, close the lid, run Fill and enter your aquarium's KH again under Tank Alkalinity (Chapter 5).

There is a video of the procedure at https://youtube.com/shorts/qm--dZ7LNUg

13.2 Memory card

The card holds the history, the event log and the interface files. When it fails, the symptoms are characteristic: graphs stop appearing, updates do not complete, or System Status shows SD: Not Detected.

The fix is to format or replace the card. Talk to support first — in many cases we can identify the cause remotely.

13.3 pH probe

  • Rinse only with drinking water — never with reverse-osmosis, distilled or deionised water, which attack the bulb. To dry it, dab it with soft paper, without rubbing the bulb.
  • Calibrate the probe every 3 months, during the chamber cleaning (section 13.1), always with the equipment empty.
  • Never let the probe dry out — more than 30 minutes in the air can damage it permanently.
  • Replace it every 12 months, or sooner if Probe Health falls below 55%.
  • When you replace the probe, log it under Configuration → pH Probe → Log probe replacement before calibrating the new one: logging clears the previous calibration.

13.4 Probe health

The health indicator is an index from 0 to 100% summarising probe quality from four factors: signal noise, slope, offset and probe age. It also appears on the home screen among the indicators.

RangeRatingWhat to do
90–100%ExcellentReliable probe — no action needed
75–89%GoodNormal operation — just watch the trend
55–74%FairRecalibrate, with the equipment empty, and plan probe replacement
below 55%PoorReplace the probe — unreliable readings

What slope is, and how to read its value, is covered in section 4.2.

13.5 Logging a probe replacement

The screen shows the installation date, how many days the probe has been in use and the recommended replacement date. When you change the probe, tap Log probe replacement to reset that count.

A new probe starts out uncalibrated: the screen shows Probe not calibrated and the warning New calibration required, and both clear on their own once you complete the wizard in section 4.2 — with the equipment empty, as in every probe calibration.